Patient’s body size: 120 – 210 cm (Infinitely variable adjustment of saddle height)
Patient’s weight: Max. 160 kg (with additional tilt protection)
Storage space: 45 x 83 cm (with additional tilt protection 81 x 83 cm)
Handlebar: 360° turnable
Control unit:: 180° turnable
Display: TFT-Display 57 x 43 mm (VGA)
(Visualization of all important ergometry data)
Supply voltage: 230-240 V with 50 Hz, 110-120 V with 60 Hz (on customer request)
Drive mechanism: Almost noiseless and maintenance-free drive unit with Poly-V belt
Braking principle: Computer controlled eddy current brake with permanent torque control (Rotation speed independent brake control)
Range of revolutions: 30 to 130 rpm, crank controlled (Independent on revolutions per minute)
Load range: 20 to 999 watts, independent on revolutions per minute (acc. standard)
Adjustable in steps of 1 watt
Load tolerance: Acc. DIN VDE 0750-238 (within RPM independent operating range)
Load precept: External by Master device (ECG, PC, …) via interface
Manual in steps of 5 watts or step profile
Time intervals: 1 to 99 minutes
Interfaces: RS232 (galvanically isolated) Bluetooth and WLAN interface to master device
Long-term accuracy: Continuous torque control and equalization according to weight